The Invacare Platinum 10L Oxygen Concentrator provides an attractive, 24-hour home oxygen solution for high flow oxygen users who are tired of inconvenient liquid or tank solutions. This Platinum 10 offers continuous flow oxygen delivery from 2 to 10 LPM. A quiet operator, the Platinum 10 is also compatible with the Invacare's Home Fill Oxygen System that enables owners to easily fill their own high-pressure tank cylinders for portable oxygen needs. The unit offers self-diagnostic electronics to minimize troubleshooting time, a top handle for ease of mobility around your home, and a battery-free power loss alarm monitor the protect the sieve from degradation. It has 3 separate filters which are reusable easy to wash and replace, to assure oxygen purity. The oxygen purity level ranges from 87-95.6%. The SensO2 indicator light will display a GREEN light during normal operation oxygen concentration levels greater than 85%, a YELLOW light for concentrations between 73-85%, and a RED light accompanied by an audible alarm for oxygen concentrations less than 73%. The compressor has a 35 PSI pressure-relief valve and a low flow alarm as well for added protection.
• Oxygen Concentrators are devices which concentrates oxygen from atmospheric air by filtering nitrogen and remaining gases present in it. • Oxygen concentrators are mainly used for patients those who have respiratory issues, or a person who need additional oxygen for a long period and home care patients. • It works based on the principle of Pressure Swing Absorption (PSA), which involves separation of gas species from a mixture of gases under suitable pressure. • Based on this principle, oxygen concentrator uses zeolite particles for separation of gases,At high pressure zeolite particles adsorbs large particles such as nitrogen and oxygen is compressed and move to other end of the valve. • This process of separation will take place in the two adjacent sieve beds between fractions of 10 seconds. These sieve beds are filled with zeolite for filtration. • One sieve bed will collect Oxygen at a time and other one will remove nitrogen through other end of the valve. It will cycled each time. • The Hepa filter and Cabinet filter used in this machine are capable of filtering large particles and minute particles up to a large extend. • On the other hand, we can say oxygen concentrators are replacement of oxygen cylinders. • Since this machine only needs electricity for working, and not need to refill all the time as it in cylinders, it is portable and because of the wheelers provided with machine, patient himself can adjust the movement if he needed. • The lifetime of this machine is more than 3-4 years, which is very much useful for patients in homecare. • Depend on the patient condition we can set the amount of oxygen supply to the patient. • There are 5L to 10 L machines available for the convenient usage of patient. • There are audible alarms indications available for proper maintenance and checking. So there is no need of care taker all the time near the patient.
Product Type |
Stationary |
Brand |
Invacare |
Model Name/Number |
Platinum 10 |
Weight (in kg) |
24.0 kg ± 1 kg |
Oxygen Flow Rate |
2 - 10 LPM |
Purity |
87%-95.6% At all flow rates |
Maximum Tubing |
2.1 m cannula with a maximum 15 m of Crush-Proof Tubing |
Sound level |
Pressure Level - < 62 dBA weighted for 3 L/min flow, < 62 dBA weighted for 10 L/min flow |
Dimensions |
36.5 x 46.7 x 67 cm ± 1 cm |
Power |
120 V AC + 10, –15% (132 V AC/102 V AC)), 60 Hz |
Power Consumption |
585 W |
Operation |
Time - Upto 24 hours per day |
Filters |
Cabinet (2), Outlet HEPA and Compressor Inlet |
Operational Temperature |
10°C to 35°C |
Operational Humidity |
1% |
Operating Altitude |
1230 m |
Storage Temperature |
-29°C to 65°C |
Storage Humidity |
Upto 95% |
Atmospheric Pressure |
101.33 kPa – 88.0 kPa |
Rated Input Current |
4.9 A |
Maximum Outlet Pressure |
62.0 kPa ± 3.45 kPa |
Pressure Relief Mechanism Operational |
35 psi ± 5 psi |
Safety System |
Current overload or line surge shutdown. |
Cabinet |
Impact Resistant flame-retardant plastic cabinet that conforms to UL 94V–0 |
Standards and Regulatory Listing |
ETL certified complying with AAMI ES60601-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, CSA C22.2#60601-1, ISO 80601-2-69 |